Description
Technical Specification :
Precooling Room
Temperature Range: 2°C to 8°C | Accuracy: ≤ ± 1.0ºC | Uniformity: ≤ ± 2.0ºC |
Freezer Room
Temperature Range: -10°C to -30°C | Accuracy: ≤ ± 3.0ºC | Uniformity: ≤ ± 5.0ºC |
Precooling Room
Internal size in mm |
External size in mm on |
Rack |
Tray |
Tray Size in mm |
Connected Load |
---|---|---|---|---|---|
2000WX1530DX200H |
2240WX2950DX2300H |
1 |
5 |
730WX500DX30H |
2.05 KW |
Freezer Room
Models |
Cap. in |
Internal size in mm |
External size in mm on |
Rack |
Tray |
Tray Size in mm |
Connected Load |
---|---|---|---|---|---|---|---|
AO/WF |
6000 |
2000WX1530DX200H |
2240WX2950DX2300H |
4 |
20 |
610WX625DX30H |
5.0 KW |
AO/WF |
8000 |
2000WX2030DX2000H |
2240WX3450DX2300H |
4 |
20 |
860WX625DX30H |
5.0 KW |
AO/WF |
10000 |
2000WX2530DX2000H |
2240WX3950DX2300H |
6 |
30 |
730WX625DX30H |
6.0 KW |
AO/WF |
12000 |
2000WX3030DX2000H |
2240WX4450DX2300H |
6 |
30 |
895WX625DX30H |
6.0 KW |
*Measured with high accuracy data-logger at sample storage area with standard Set Point
*Custom built Chambers are available in any dimensions and capacities.
Standard Chamber Features & Provisions:
Construction: PCGI/SS 304/SS 316 Tongue & Groove Type Panels with PUF density of 40Kg/m³-Thickness: 120 mm for Deep Freezer With Pre-cooling Chamber (Ante-Room): Panel Thickness 80 mm
Sample Storage:
- Racks: Heavy Duty SS 304 Frame secured strongly to chamber walls for strength.
- Trays: Perforated types (CNC punched). Strengthened with support rails to prevent deformations.
Door 1: For Ante Chamber: PUF Insulated door with Viewing window of Double Layered Toughened Glass, mounted with imported hinges & locks.
Door 2: For Deep Freezer: PUF Insulated door mounted with imported hinges & locks.
Door Operation: Passcode protected door – from software & HMI.
Control System:
- Ergonomically designed control panel.
- Ferrule printed wiring.
- Advanced PLC Program with built in processor-based Auto Tuning.
- 2 Separate 4.3″ HMI Color LED Display. (For both: Ante Room & Deep Freezer)
- Built in Redundancy & Safety.
- Advanced High Accuracy “Transmitter Sensors” with output of 4-20 mA.
Temperature Control:
- Efficient, Reliable, Time tested & Trustworthy Refrigeration Design. Frost Free with De-frosting system.
- Built in redundancy with PLC based Auto Change Over during power outage or component failure
Safety:
- PLC Safety Cut Off for Temperature: Low, Lower & High safety cut off.
- Human Safety: “Person Trap” button provided inside the Ante chamber, which triggers Audio Visual alarm. The button enables the trapped person to
- come out even if the door is physically locked from outside
- Emergency Stop: Button provided from outside.