Description
Technical Specification :
Temperature Range: 10°C above ambient to 60°C | Accuracy :< ± 0.2°C | Uniformity :< ± 1.0°C* |
Models |
Cap. in |
Internal size in mm |
External size in mm on |
Tray |
Tray Size in mm |
Connected Load |
---|---|---|---|---|---|---|
AO/1-15 |
150 |
500Wx400Dx750H |
625Wx860Dx1350H |
2 |
405Wx400Dx30H |
1.25 KW |
AO/1-25 |
250 |
700Wx400D×900H |
825Wx860Dx1500H |
2 |
605Wx400Dx30H |
1.25 KW |
AO/1-40 | 400 |
700Wx500Dx1150H |
825Wx960Dx1750H |
3 |
605Wx500Dx30H |
1.25 KW |
AO/1-50 |
500 |
700Wx600Dx1200H |
825Wx1060Dx1800H |
3 |
605Wx600Dx30H |
1.25 KW |
AO/1-60 |
600 |
700Wx600Dx1450H |
825Wx1060Dx2050H |
3 |
605Wx600Dx30H |
1.25 KW |
AO/1-80 |
800 |
700Wx800Dx1450H |
825Wx1260Dx2050H |
4 |
605Wx800Dx30H |
1.5 KW |
AO/1-100 |
1000 |
830Wx830Dx1450H |
955Wx1290Dx2050H |
4 |
735Wx830Dx30H |
1.5 KW |
*Measured with high accuracy data-logger at sample storage area with standard Set Point
Standard Chamber Features & Provisions:
Construction: Dynamically Designed PUF filled Double walled, Inner Tank & outer body with SS 304 PUF Density of 40 kg/m³ Thickness 60mm.
Sample Storage: Trays: Perforated types (CNC punched). Strengthened with support rails to prevent deformations.
Door: PUF Insulated External door mounted with imported hinges & locks, Inner glass doors with latches. Door frame is secured with a silicone rubber gasket on the entire perimeter to ensure an air tight, leak tight chamber.
Door Operation: Biometric &/or Password protected-from software & HMI
Control System:
- Ergonomically designed Control Panel.
- Ferrule printed wiring.
- Advanced PLC Program with built in processor-based Auto Tuning.
- 4.3″ HMI Color LCD Display.
- Built in Redundancy & Safety.
- Advanced High Accuracy “Transmitter Sensors” with output of 4-20 mA.
Temperature Control:
- Efficient, Reliable, Time tested & Trustworthy Refrigeration Design.
- Built in redundancy with PLC based Auto Change Over during power outage or component failure (Compressor)
- PLC Controlled “U” Shaped SS Air Heater Banks.
Optional Backup System: PLC based Auto Changeover Backup system for Refrigeration
Safety: Dedicated Thermostats for all key points: Chamber High Temperature, Utility High Temperature.
PLC Safety Cut Off for Temperature: High, Higher & Low, Lower safety cut off