Description
Technical Specification :
Temperature Range: 2°C to 8°C | Accuracy: ≤ +/- 1.0°C | Uniformity: ≤ +/- 2.0°C* |
Models |
Cap. in |
Internal size in mm |
External size in mm on |
Tray |
Tray Size in mm |
Connected Load |
---|---|---|---|---|---|---|
AO/CC-15 |
150 |
500Wx400Dx750H |
625Wx860Dx1350H |
2 |
405Wx400Dx30H |
0.75 KW |
AO/CC-25 |
250 |
700Wx400D×900H |
825Wx860Dx1500H |
2 |
605Wx400Dx30H |
0.75 KW |
AO/CC-40 | 400 |
700Wx500Dx1150H |
825Wx960Dx1750H |
3 |
605Wx500Dx30H |
0.75 KW |
AO/CC-50 |
500 |
700Wx600Dx1200H |
825Wx1060Dx1800H |
3 |
605Wx600Dx30H |
0.75 KW |
AO/CC-60 |
600 |
700Wx600Dx1450H |
825Wx1060Dx2050H |
3 |
605Wx600Dx30H |
0.75 KW |
AO/CC-80 |
800 |
700Wx800Dx1450H |
825Wx1260Dx2050H |
4 |
605Wx800Dx30H |
1 KW |
AO/CC-100 |
1000 |
830Wx830Dx1450H |
955Wx1290Dx2050H |
4 |
735Wx830Dx30H |
1 KW |
*Measured with high accuracy data-logger at sample storage area with standard Set Point
Standard Chamber Features & Provisions:
Construction: Dynamically Designed PUF filled Double walled, Inner Tank & outer body with PCGI/SS 304/SS 316 with PUF Density of 40Kg/m³ – Thickness 60mm.
Sample Storage: Trays: Perforated types (CNC punched). Strengthened with support rails to prevent deformations.
Door: PUF Insulated External door mounted with imported hinges & locks, Inner glass doors with latches. Door frame is secured with a silicone gasket on the entire perimeter to ensure an air tight, leak tight chamber.
Door Operation: Passcode protected door operation from software & HMI
Control System:
- Ergonomically designed control panel.
- Ferrule printed wiring.
- Advanced PLC Program with built in processor-based Auto Tuning.
- 4.3″ HMI Color LED Display.
- Built in Redundancy & Safety.
- Advanced High Accuracy “Transmitter based sensors” with output of 4-20 mA.
Temperature Control:
- Efficient, Reliable, Time tested & Trustworthy Refrigeration Design.
- Built in redundancy with PLC based Auto Change Over during power outage or component failure (Compressor)
Optional Backup System: PLC based Auto Changeover Backup system for Refrigeration
Safety: PLC based Safety Cut Off for Temperature: Low, Lower & High safety cut off.