Description
Technical Specification :
Temperature Range: 20°C to 50°C | Accuracy: < ± 0.2°C | Uniformity: < ± 1.0°C* |
Rel. Humidity Range: 40%RH to 85% | RH Accuracy : < ± 1% | RH Uniformity: < ±3%RH* |
Models |
Cap. in |
Internal size in mm |
External size in mm on |
Tray |
Tray Size in mm |
Connected Load |
---|---|---|---|---|---|---|
AO/PSC-20 |
200 |
700Wx500Dx600H |
915Wx1155Dx1670H |
2 |
605Wx500Dx30H |
3.4 KW |
*Measured with high accuracy data-logger at sample storage area with standard Set Point
Standard Chamber Features & Provisions:
Construction: Dynamically Designed PUF filled Double walled, Inner Tank & outer body with SS 304, PUF Density 40kg/m³ with 60 mm thickness
Sample Storage: SS Trays.
Door: PUF Insulated External door mounted with imported hinges & locks, Inner glass doors with latches. Door frame is secured with a silicone rubber gasket on the entire perimeter to ensure an air tight, leak tight chamber.
Door Operation: Biometric &/or Password protected-from software & HMI
UV & Fluorescent Lights:
- As per ICH Guidelines-1.2 Million Lux Hours of visible light and 200 watt Hr/m2 of UV light. Auto shut Off after specified exposure level of Lux & UV Intensity.
- Individual Microprocessor based device is provided to start exposure of UV and Fluorescent lights as required.
- UV light is switched off automatically once the inner glass door is opened to prevent hazardous exposure: for Human safety.
- Samples can be tested under Fluorescent light or U.V. lights separately as well as combine as per test requirement.
Optional System:
- Allyone Photostability Chambers can be with Temperature or with Temp & RH (Humidification)
- Lux meter, UV Meter
- Printer Interface (for measuring Lux & UV intensity)
Control System:
- Ergonomically designed Control Panel.
- Ferrule printed wiring.
- Advanced PLC Program with built in processor-based Auto Tuning.
- 4.3″ HMI Color LED Display.
- Built in Redundancy & Safety.
- Advanced High Accuracy “Transmitter Sensors” with output of 4-20 mA.
Temperature Control:
- Efficient, Reliable, Time tested & Trustworthy Refrigeration Design.
- PLC Controlled “U” Shaped SS Air Heater Banks.
Humidity Control: S S Reservoirs- steam kettles form a Humidification system.
Safety: Dedicated Thermostats for all key points: Chamber High Temperature, Steam Kettle Element, Utility High Temperature.
PLC Safety Cut Off for Temperature & Humidity: High, Higher & Low, Lower safety cut off