Description
Technical Specification :
ModTemperature Range: 20°C to 60°C el |
Accuracy :< ± 0.2°C |
Uniformity :< ± 1.0°C* |
Rel. Humidity Range: 40%R H to 90%RH |
Accuracy :< ± 1.0% |
Uniformity :< ± 3.0%* |
Standard Equipment Operating Conditions
- 25°C / 60%RH
- 30°C / 65%RH
- 30°C / 75%RH
- 40°C / 75%RH
Low Humidity Equipment Operating Conditions
- 25°C / 40%RH
- 30°C / 35%RH
- 40°C / NMT25%RH
Models |
Cap. in |
Internal size in mm |
External size in mm on |
Rack |
Tray |
Tray Size in mm |
Connected Load |
---|---|---|---|---|---|---|---|
AO/WSC-400 |
4000 |
2000WX1000DX2000H |
2160WX1630DX2200H |
2 |
10 |
730WX625DX30H |
4.0 KW |
AO/WSC-600 |
6000 |
2000WX1500DX2000H |
2160WX2130DX2200H |
4 |
20 |
600WX625DX30H |
4.25 KW |
AO/WSC-800 |
8000 |
2000WX2000DX2000H |
2160WX2630DX2200H |
4 |
20 |
850WX625DX30H |
4.25 KW |
AO/WSC-1000 |
10000 |
2000WX2500DX2000H |
2160WX3130DX2200H |
6 |
30 |
720WX625DX30H |
4.25 KW |
AO/WSC-1200 |
12000 |
2000WX3000DX2000H |
2160WX3630DX2200H |
6 |
30 |
885WX625DX30H |
4.25 KW |
AO/WSC-1600 |
16000 |
2000WX4000DX2000H |
2160WX4630DX2200H |
8 |
40 |
905WX625DX30H |
5.5 KW |
AO/WSC-2000 |
20000 |
2000WX4000DX2500H |
2160WX4630DX2700H |
8 |
48 |
905WX625DX30H |
6.0 KW |
AO/WSC-2500 |
25000 |
2300WX4500DX2500H |
2460WX5130DX2700H |
10 |
60 |
815WX750DX30H |
6.0 KW |
AO/WSC-3000 |
30000 |
4000WX3000DX2500H |
4160W X3630DX2700H |
12 |
72 |
885WX625DX30H |
9.0 KW |
AO/WSC-4000 |
40000 |
4000W X4000DX2500H |
4160W X4630DX2700H |
16 |
96 |
905WX625DX30H |
9.0 KW |
AO/WSC-5000 |
50000 |
4500W X4500DX2500H |
4660W X5130DX2700H |
20 |
120 |
815WX725DX30H |
12.5 KW |
*Measured with high accuracy data-logger at sample storage area with standard Set Point
*Custom built Chambers are available in any dimensions and capacities.
Standard Chamber Features & Provisions:
Construction: PCGI/SS 304/SS 316 Tongue & Groove Type Panels with PUF Density of 40Kg/m3 with thickness of 80 mm.
Sample Storage:
- Racks: Heavy Duty SS 304 frame secured firmly to chamber inner walls for strength.
- Trays: Perforated types (CNC punched). Strengthened with support rails to prevent deformations.
Door: PUF Insulated door with Full View double layered Toughened Glass, mounted with imported hinges & locks.
Door Operation: Passcode protected door operation — from software & HMI
Control System:
- Ergonomically designed control panel.
- Ferrule printed wiring.
- Advanced PLC program with built in processor-based Auto Tuning.
- 3″ HMI Color LCD Display.
- Built in Redundancy & Safety.
- Advanced High Accuracy “Transmitter based sensors” with output of 4-20 mA.
Temperature Control:
- Efficient, Reliable, Time tested & Trustworthy refrigeration design.
- Built in redundancy with PLC based Auto Change Over during power outage or component failure viz: Compressor.
- PLC Controlled “U” Shaped SS Air Heater Banks.
Safety: Dedicated Thermostats for all key points: Chamber High Temperature, Utility High Temperature
PLC Safety Cut Off for Temperature: High, Higher & Low, Lower safety cut off .
Human Safety: “Person Trap” button provided inside the chamber, which triggers Audio Visual alarm. The button enables the trapped person to come out even if the door is physically locked from outside
Emergency Stop: button provided from outside.
Walk-In Humidity Chamber
Stability testing is of paramount and inseparable part when it comes to assessing the Assurance of Quality (QA)& Controlling the Quality (QC) of pharmaceutical Drug substance, formulations and products.
Testing stability of New Drugs, Formulations & Pharmaceutical Substances is one of the most important prerequisites in the licencing process those are defined by various Regulatory Authorities of the world like ICH (Guideline Q1 AR2), US FDA (21 CFR Part 11), PICS, MHRA, Health Canada, TGA, WHO, ANVISA, ISPE & so on
These Regulatory Authorities defines and harmonize the various stability Test conditions that are to be performed in order to provide authentic & verified evidence of the stability of the formulations.
Data from these Stability studies enable recommendation & establishment of storage conditions, retest intervals and shelf lives: projected shelf life & expiration and most importantly for New Product Development
The shelf life of a product is commonly estimated using types of stability testing: Real-time, Intermediate, Long Term & Accelerated.
In real-time stability testing, a product is stored at recommended storage conditions and monitored until it fails the specification.
In accelerated stability tests, a product is stored at elevated stress conditions of high temperature & humidity.
ALLYONE mastered & perfected the art & science of manufacturing Stability (Humidity) Chambers: an ability to maintain a stable temperature, and a stable value of relative humidity. We have a wide range of chambers for you to choose from as per your requirement: Standard & Customized.
Our Stability (Humidity) Chambers have a reputation to quickly achieve & maintain the defined setpoint to simulate the condition perfectly & flawlessly with extraordinary control on both Temperature & Humidity along with outstanding uniformity.
Our Stability (Humidity) Chambers are conceptualized, meticulously designed and are built by artisans & professionals which meets and encompasses all
the statutory regulatory requirements laid by world’s leading Regulatory Authorities & can be used to perform Stability Studies defined by ICH Guideline Q1AR2 for Real, Intermediate, Long Term & Accelerated types.
We have gained mastery in designing Specially designed chambers for “Low Temperature-Low Humidity”, “High Temperature, Low Humidity” operating conditions as well.
ALLYONE Stability Chambers comes with a dynamic, User Friendly, Future ready, most advanced proprietary software “ALLYSOFT INTE GRO 2.0” compatible with US FDA 21 CFR Part 11 norms offering complete control & freedom in the hands of user so that the user can focus on the more important tasks on his/her hands!
“ALLYSOFT INTE GRO 2.0” software precisely orchestrates all the utilities of the machine to achieve the Set Point at the quickest and maintains the condition even in the most demanding conditions without compromising on performance. The software records the test data accurately at all times without missing a single bit.
Extra emphasis has been given on to the safety of users and equipment by providing multiple “Safety Features” in form of Audio-Visual Alarms ensuring trouble free performance & extended life of machine.
ALLYONE Stability (Humidity) Chambers comes with the complete statutory Qualification & Validation Documentation part as laid down by cGMP Norms & ISPE recommendations which is of paramount importance in Pharmaceutical & Allied Industries: Design Qualification (DQ), Installation Qualification (IQ), Operational Qualification (OQ) & Performance Qualification (PQ) of the equipment, as well as Operational Qualification (OQ) & Performance Qualification (PQ) for the software “ALLYSOFT INTE GRO 2.0”